Reducing SAR in parallel excitation using variable-density spirals

a simulation-based study

Yinan Liu, Ke Feng, Mary P. McDougall, Steven M. Wright, Jim Ji

Research output: Contribution to journalArticle

14 Citations (Scopus)

Abstract

Parallel excitation using multiple transmit channels has emerged as an effective method to shorten multidimensional spatially selective radiofrequency (RF) pulses, which have a number of important applications, including B1 field inhomogeneity correction in high-field MRI. The specific absorption rate (SAR) is a primary concern in high-field MRI, where wavelength effects can lead to local peaks in SAR. In parallel excitation, the subjects are exposed to RF pulses from multiple coils, which makes the SAR problem more complex to analyze, yet potentially enables greater freedom in designing RF pulses with lower SAR. Parallel-excitation techniques typically employ either Cartesian or constant-density (CD) spiral trajectories. In this article, variable-density (VD) spiral trajectories are explored as a means for SAR reduction in parallel-excitation pulse design. Numerical simulations were conducted to study the effects of CD and VD spirals on parallel excitation. Specifically, the electromagnetic fields of a four-channel transmit head coil with a three-dimensional head model at 4.7 T were simulated using a finite-difference time domain method. The parallel RF pulses were designed and the resulting excitation patterns were generated using a Bloch simulator. The SAR distributions due to CD and VD spirals were evaluated quantitatively. The simulation results show that, for the same pulse duration, parallel excitation with VD spirals can achieve a lower SAR compared to CD spirals for parallel excitation. VD spirals also resulted in reduced artifact power in the excitation patterns. This gain came with slight, but noticeable, degrading of the spatial resolution of the resulting excitation patterns.

Original languageEnglish
Pages (from-to)1122-1132
Number of pages11
JournalMagnetic Resonance Imaging
Volume26
Issue number8
DOIs
Publication statusPublished - Oct 2008
Externally publishedYes

Fingerprint

Magnetic resonance imaging
Trajectories
Finite difference time domain method
Electromagnetic fields
excitation
simulation
Simulators
Wavelength
Computer simulation
pulses
Head
Electromagnetic Fields
coils
trajectories
Artifacts
finite difference time domain method
simulators
artifacts
electromagnetic fields
pulse duration

Keywords

  • Parallel excitation
  • SAR
  • Transmit SENSE
  • Variable-density spiral

ASJC Scopus subject areas

  • Biophysics
  • Clinical Biochemistry
  • Structural Biology
  • Radiology Nuclear Medicine and imaging
  • Condensed Matter Physics

Cite this

Reducing SAR in parallel excitation using variable-density spirals : a simulation-based study. / Liu, Yinan; Feng, Ke; McDougall, Mary P.; Wright, Steven M.; Ji, Jim.

In: Magnetic Resonance Imaging, Vol. 26, No. 8, 10.2008, p. 1122-1132.

Research output: Contribution to journalArticle

Liu, Yinan ; Feng, Ke ; McDougall, Mary P. ; Wright, Steven M. ; Ji, Jim. / Reducing SAR in parallel excitation using variable-density spirals : a simulation-based study. In: Magnetic Resonance Imaging. 2008 ; Vol. 26, No. 8. pp. 1122-1132.
@article{911a268e9af04e1e9a2072a09f390688,
title = "Reducing SAR in parallel excitation using variable-density spirals: a simulation-based study",
abstract = "Parallel excitation using multiple transmit channels has emerged as an effective method to shorten multidimensional spatially selective radiofrequency (RF) pulses, which have a number of important applications, including B1 field inhomogeneity correction in high-field MRI. The specific absorption rate (SAR) is a primary concern in high-field MRI, where wavelength effects can lead to local peaks in SAR. In parallel excitation, the subjects are exposed to RF pulses from multiple coils, which makes the SAR problem more complex to analyze, yet potentially enables greater freedom in designing RF pulses with lower SAR. Parallel-excitation techniques typically employ either Cartesian or constant-density (CD) spiral trajectories. In this article, variable-density (VD) spiral trajectories are explored as a means for SAR reduction in parallel-excitation pulse design. Numerical simulations were conducted to study the effects of CD and VD spirals on parallel excitation. Specifically, the electromagnetic fields of a four-channel transmit head coil with a three-dimensional head model at 4.7 T were simulated using a finite-difference time domain method. The parallel RF pulses were designed and the resulting excitation patterns were generated using a Bloch simulator. The SAR distributions due to CD and VD spirals were evaluated quantitatively. The simulation results show that, for the same pulse duration, parallel excitation with VD spirals can achieve a lower SAR compared to CD spirals for parallel excitation. VD spirals also resulted in reduced artifact power in the excitation patterns. This gain came with slight, but noticeable, degrading of the spatial resolution of the resulting excitation patterns.",
keywords = "Parallel excitation, SAR, Transmit SENSE, Variable-density spiral",
author = "Yinan Liu and Ke Feng and McDougall, {Mary P.} and Wright, {Steven M.} and Jim Ji",
year = "2008",
month = "10",
doi = "10.1016/j.mri.2008.02.003",
language = "English",
volume = "26",
pages = "1122--1132",
journal = "Magnetic Resonance Imaging",
issn = "0730-725X",
publisher = "Elsevier Inc.",
number = "8",

}

TY - JOUR

T1 - Reducing SAR in parallel excitation using variable-density spirals

T2 - a simulation-based study

AU - Liu, Yinan

AU - Feng, Ke

AU - McDougall, Mary P.

AU - Wright, Steven M.

AU - Ji, Jim

PY - 2008/10

Y1 - 2008/10

N2 - Parallel excitation using multiple transmit channels has emerged as an effective method to shorten multidimensional spatially selective radiofrequency (RF) pulses, which have a number of important applications, including B1 field inhomogeneity correction in high-field MRI. The specific absorption rate (SAR) is a primary concern in high-field MRI, where wavelength effects can lead to local peaks in SAR. In parallel excitation, the subjects are exposed to RF pulses from multiple coils, which makes the SAR problem more complex to analyze, yet potentially enables greater freedom in designing RF pulses with lower SAR. Parallel-excitation techniques typically employ either Cartesian or constant-density (CD) spiral trajectories. In this article, variable-density (VD) spiral trajectories are explored as a means for SAR reduction in parallel-excitation pulse design. Numerical simulations were conducted to study the effects of CD and VD spirals on parallel excitation. Specifically, the electromagnetic fields of a four-channel transmit head coil with a three-dimensional head model at 4.7 T were simulated using a finite-difference time domain method. The parallel RF pulses were designed and the resulting excitation patterns were generated using a Bloch simulator. The SAR distributions due to CD and VD spirals were evaluated quantitatively. The simulation results show that, for the same pulse duration, parallel excitation with VD spirals can achieve a lower SAR compared to CD spirals for parallel excitation. VD spirals also resulted in reduced artifact power in the excitation patterns. This gain came with slight, but noticeable, degrading of the spatial resolution of the resulting excitation patterns.

AB - Parallel excitation using multiple transmit channels has emerged as an effective method to shorten multidimensional spatially selective radiofrequency (RF) pulses, which have a number of important applications, including B1 field inhomogeneity correction in high-field MRI. The specific absorption rate (SAR) is a primary concern in high-field MRI, where wavelength effects can lead to local peaks in SAR. In parallel excitation, the subjects are exposed to RF pulses from multiple coils, which makes the SAR problem more complex to analyze, yet potentially enables greater freedom in designing RF pulses with lower SAR. Parallel-excitation techniques typically employ either Cartesian or constant-density (CD) spiral trajectories. In this article, variable-density (VD) spiral trajectories are explored as a means for SAR reduction in parallel-excitation pulse design. Numerical simulations were conducted to study the effects of CD and VD spirals on parallel excitation. Specifically, the electromagnetic fields of a four-channel transmit head coil with a three-dimensional head model at 4.7 T were simulated using a finite-difference time domain method. The parallel RF pulses were designed and the resulting excitation patterns were generated using a Bloch simulator. The SAR distributions due to CD and VD spirals were evaluated quantitatively. The simulation results show that, for the same pulse duration, parallel excitation with VD spirals can achieve a lower SAR compared to CD spirals for parallel excitation. VD spirals also resulted in reduced artifact power in the excitation patterns. This gain came with slight, but noticeable, degrading of the spatial resolution of the resulting excitation patterns.

KW - Parallel excitation

KW - SAR

KW - Transmit SENSE

KW - Variable-density spiral

UR - http://www.scopus.com/inward/record.url?scp=51649096117&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=51649096117&partnerID=8YFLogxK

U2 - 10.1016/j.mri.2008.02.003

DO - 10.1016/j.mri.2008.02.003

M3 - Article

VL - 26

SP - 1122

EP - 1132

JO - Magnetic Resonance Imaging

JF - Magnetic Resonance Imaging

SN - 0730-725X

IS - 8

ER -