Experimental limitations in impedance spectroscopy: Part II - Electrode artifacts in three-point measurements on Pt/YSZ

G. Hsieh, T. O. Mason, L. R. Pederson

    Research output: Contribution to journalArticle

    37 Citations (Scopus)

    Abstract

    High temperature (570-750°C) three-point impedance measurements on the system Pt/YSZ using a platinum reference electrode exhibited distortions/artifacts due to the voltage divider effect. These distortions could be avoided by using a lower impedance reference electrode (Ag) or employing an instrument with a higher input impedance. A correction procedure was also successful.