Chemical exchange saturation transfer magnetic resonance imaging of human knee cartilage at 3 T and 7 T

Anup Singh, Mohammad Haris, Kejia Cai, Victor B. Kassey, Feliks Kogan, Damodar Reddy, Hari Hariharan, Ravinder Reddy

Research output: Contribution to journalArticle

97 Citations (Scopus)

Abstract

The sensitivity of chemical exchange saturation transfer (CEST) on glycosaminoglycans (GAGs) in human knee cartilage (gagCEST) in vivo was evaluated at 3 and 7 T field strengths. Calculated gagCEST values without accounting for B0 inhomogeneity (∼0.6 ppm) were >20%. After B0 inhomogeneity correction, calculated gagCEST values were negligible at 3 T and ∼6% at 7 T. These results suggest that accurate B 0 correction is a prerequisite for observing reliable gagCEST. Results obtained with varying saturation pulse durations and amplitudes as well as the consistency between numerical simulations and our experimental results indicate that the negligible gagCEST observed at 3 T is due to direct saturation effects and fast exchange rate. As GAG loss from cartilage is expected to result in a further reduction in gagCEST, gagCEST method is not expected to be clinically useful at 3 T. At high fields such as 7 T, this method holds promise as a viable clinical technique. Magn Reson Med, 2012. © 2011 Wiley Periodicals, Inc.

Original languageEnglish
Pages (from-to)588-594
Number of pages7
JournalMagnetic Resonance in Medicine
Volume68
Issue number2
DOIs
Publication statusPublished - Aug 2012
Externally publishedYes

Fingerprint

Glycosaminoglycans
Cartilage
Knee
Magnetic Resonance Imaging

Keywords

  • cartilage
  • CEST
  • GAG
  • gagCEST
  • MRI
  • saturation pulse

ASJC Scopus subject areas

  • Radiology Nuclear Medicine and imaging

Cite this

Chemical exchange saturation transfer magnetic resonance imaging of human knee cartilage at 3 T and 7 T. / Singh, Anup; Haris, Mohammad; Cai, Kejia; Kassey, Victor B.; Kogan, Feliks; Reddy, Damodar; Hariharan, Hari; Reddy, Ravinder.

In: Magnetic Resonance in Medicine, Vol. 68, No. 2, 08.2012, p. 588-594.

Research output: Contribution to journalArticle

Singh, Anup ; Haris, Mohammad ; Cai, Kejia ; Kassey, Victor B. ; Kogan, Feliks ; Reddy, Damodar ; Hariharan, Hari ; Reddy, Ravinder. / Chemical exchange saturation transfer magnetic resonance imaging of human knee cartilage at 3 T and 7 T. In: Magnetic Resonance in Medicine. 2012 ; Vol. 68, No. 2. pp. 588-594.
@article{ad6e6f4d97134b9a90712e8a974bff9b,
title = "Chemical exchange saturation transfer magnetic resonance imaging of human knee cartilage at 3 T and 7 T",
abstract = "The sensitivity of chemical exchange saturation transfer (CEST) on glycosaminoglycans (GAGs) in human knee cartilage (gagCEST) in vivo was evaluated at 3 and 7 T field strengths. Calculated gagCEST values without accounting for B0 inhomogeneity (∼0.6 ppm) were >20{\%}. After B0 inhomogeneity correction, calculated gagCEST values were negligible at 3 T and ∼6{\%} at 7 T. These results suggest that accurate B 0 correction is a prerequisite for observing reliable gagCEST. Results obtained with varying saturation pulse durations and amplitudes as well as the consistency between numerical simulations and our experimental results indicate that the negligible gagCEST observed at 3 T is due to direct saturation effects and fast exchange rate. As GAG loss from cartilage is expected to result in a further reduction in gagCEST, gagCEST method is not expected to be clinically useful at 3 T. At high fields such as 7 T, this method holds promise as a viable clinical technique. Magn Reson Med, 2012. {\circledC} 2011 Wiley Periodicals, Inc.",
keywords = "cartilage, CEST, GAG, gagCEST, MRI, saturation pulse",
author = "Anup Singh and Mohammad Haris and Kejia Cai and Kassey, {Victor B.} and Feliks Kogan and Damodar Reddy and Hari Hariharan and Ravinder Reddy",
year = "2012",
month = "8",
doi = "10.1002/mrm.23250",
language = "English",
volume = "68",
pages = "588--594",
journal = "Magnetic Resonance in Medicine",
issn = "0740-3194",
publisher = "John Wiley and Sons Inc.",
number = "2",

}

TY - JOUR

T1 - Chemical exchange saturation transfer magnetic resonance imaging of human knee cartilage at 3 T and 7 T

AU - Singh, Anup

AU - Haris, Mohammad

AU - Cai, Kejia

AU - Kassey, Victor B.

AU - Kogan, Feliks

AU - Reddy, Damodar

AU - Hariharan, Hari

AU - Reddy, Ravinder

PY - 2012/8

Y1 - 2012/8

N2 - The sensitivity of chemical exchange saturation transfer (CEST) on glycosaminoglycans (GAGs) in human knee cartilage (gagCEST) in vivo was evaluated at 3 and 7 T field strengths. Calculated gagCEST values without accounting for B0 inhomogeneity (∼0.6 ppm) were >20%. After B0 inhomogeneity correction, calculated gagCEST values were negligible at 3 T and ∼6% at 7 T. These results suggest that accurate B 0 correction is a prerequisite for observing reliable gagCEST. Results obtained with varying saturation pulse durations and amplitudes as well as the consistency between numerical simulations and our experimental results indicate that the negligible gagCEST observed at 3 T is due to direct saturation effects and fast exchange rate. As GAG loss from cartilage is expected to result in a further reduction in gagCEST, gagCEST method is not expected to be clinically useful at 3 T. At high fields such as 7 T, this method holds promise as a viable clinical technique. Magn Reson Med, 2012. © 2011 Wiley Periodicals, Inc.

AB - The sensitivity of chemical exchange saturation transfer (CEST) on glycosaminoglycans (GAGs) in human knee cartilage (gagCEST) in vivo was evaluated at 3 and 7 T field strengths. Calculated gagCEST values without accounting for B0 inhomogeneity (∼0.6 ppm) were >20%. After B0 inhomogeneity correction, calculated gagCEST values were negligible at 3 T and ∼6% at 7 T. These results suggest that accurate B 0 correction is a prerequisite for observing reliable gagCEST. Results obtained with varying saturation pulse durations and amplitudes as well as the consistency between numerical simulations and our experimental results indicate that the negligible gagCEST observed at 3 T is due to direct saturation effects and fast exchange rate. As GAG loss from cartilage is expected to result in a further reduction in gagCEST, gagCEST method is not expected to be clinically useful at 3 T. At high fields such as 7 T, this method holds promise as a viable clinical technique. Magn Reson Med, 2012. © 2011 Wiley Periodicals, Inc.

KW - cartilage

KW - CEST

KW - GAG

KW - gagCEST

KW - MRI

KW - saturation pulse

UR - http://www.scopus.com/inward/record.url?scp=84863827545&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84863827545&partnerID=8YFLogxK

U2 - 10.1002/mrm.23250

DO - 10.1002/mrm.23250

M3 - Article

C2 - 22213239

AN - SCOPUS:84863827545

VL - 68

SP - 588

EP - 594

JO - Magnetic Resonance in Medicine

JF - Magnetic Resonance in Medicine

SN - 0740-3194

IS - 2

ER -