A pocket-sized disposable device for testing the integrity of sensation in the outpatient setting

F. L. Bowling, C. A. Abbott, W. E. Harris, S. Atanasov, Rayaz Malik, A. J M Boulton

Research output: Contribution to journalArticle

23 Citations (Scopus)

Abstract

Aims To compare the Ipswich Touch Test and the VibraTip with the Neuropathy Disability Score and the vibration perception threshold for detecting the 'at-risk' foot. Methods We directly compared the Ipswich Touch Test and the VibraTip with both the Neuropathy Disability Score ≥6 and the vibration perception threshold ≥25V indicating 'at-risk' feet in 83 individuals. Results The vibration perception threshold and Neuropathy Disability Score tests exhibited almost perfect agreement with each other (P<0.001). The VibraTip and Ipswich Touch Test results were identical (P<0.001). The VibraTip and Ipswich Touch Test results also exhibited almost perfect agreement with the vibration perception threshold (P<0.001) and the Neuropathy Disability Score (P<0.001). Conclusions These two simple and efficient tests are easy to teach, reliable and can be used in any setting, and neither requires an external power source. We conclude that both the VibraTip and the Ipswich Touch Test are reliable and sensitive tests for identifying the 'high-risk' foot.

Original languageEnglish
Pages (from-to)1550-1552
Number of pages3
JournalDiabetic Medicine
Volume29
Issue number12
DOIs
Publication statusPublished - Dec 2012
Externally publishedYes

Fingerprint

Touch
Vibration
Outpatients
Equipment and Supplies
Foot
Electric Power Supplies

ASJC Scopus subject areas

  • Endocrinology
  • Internal Medicine
  • Endocrinology, Diabetes and Metabolism

Cite this

A pocket-sized disposable device for testing the integrity of sensation in the outpatient setting. / Bowling, F. L.; Abbott, C. A.; Harris, W. E.; Atanasov, S.; Malik, Rayaz; Boulton, A. J M.

In: Diabetic Medicine, Vol. 29, No. 12, 12.2012, p. 1550-1552.

Research output: Contribution to journalArticle

Bowling, F. L. ; Abbott, C. A. ; Harris, W. E. ; Atanasov, S. ; Malik, Rayaz ; Boulton, A. J M. / A pocket-sized disposable device for testing the integrity of sensation in the outpatient setting. In: Diabetic Medicine. 2012 ; Vol. 29, No. 12. pp. 1550-1552.
@article{fa72f6ccd23b4acc9e27114a1c715c1e,
title = "A pocket-sized disposable device for testing the integrity of sensation in the outpatient setting",
abstract = "Aims To compare the Ipswich Touch Test and the VibraTip with the Neuropathy Disability Score and the vibration perception threshold for detecting the 'at-risk' foot. Methods We directly compared the Ipswich Touch Test and the VibraTip with both the Neuropathy Disability Score ≥6 and the vibration perception threshold ≥25V indicating 'at-risk' feet in 83 individuals. Results The vibration perception threshold and Neuropathy Disability Score tests exhibited almost perfect agreement with each other (P<0.001). The VibraTip and Ipswich Touch Test results were identical (P<0.001). The VibraTip and Ipswich Touch Test results also exhibited almost perfect agreement with the vibration perception threshold (P<0.001) and the Neuropathy Disability Score (P<0.001). Conclusions These two simple and efficient tests are easy to teach, reliable and can be used in any setting, and neither requires an external power source. We conclude that both the VibraTip and the Ipswich Touch Test are reliable and sensitive tests for identifying the 'high-risk' foot.",
author = "Bowling, {F. L.} and Abbott, {C. A.} and Harris, {W. E.} and S. Atanasov and Rayaz Malik and Boulton, {A. J M}",
year = "2012",
month = "12",
doi = "10.1111/j.1464-5491.2012.03730.x",
language = "English",
volume = "29",
pages = "1550--1552",
journal = "Diabetic Medicine",
issn = "0742-3071",
publisher = "Wiley-Blackwell",
number = "12",

}

TY - JOUR

T1 - A pocket-sized disposable device for testing the integrity of sensation in the outpatient setting

AU - Bowling, F. L.

AU - Abbott, C. A.

AU - Harris, W. E.

AU - Atanasov, S.

AU - Malik, Rayaz

AU - Boulton, A. J M

PY - 2012/12

Y1 - 2012/12

N2 - Aims To compare the Ipswich Touch Test and the VibraTip with the Neuropathy Disability Score and the vibration perception threshold for detecting the 'at-risk' foot. Methods We directly compared the Ipswich Touch Test and the VibraTip with both the Neuropathy Disability Score ≥6 and the vibration perception threshold ≥25V indicating 'at-risk' feet in 83 individuals. Results The vibration perception threshold and Neuropathy Disability Score tests exhibited almost perfect agreement with each other (P<0.001). The VibraTip and Ipswich Touch Test results were identical (P<0.001). The VibraTip and Ipswich Touch Test results also exhibited almost perfect agreement with the vibration perception threshold (P<0.001) and the Neuropathy Disability Score (P<0.001). Conclusions These two simple and efficient tests are easy to teach, reliable and can be used in any setting, and neither requires an external power source. We conclude that both the VibraTip and the Ipswich Touch Test are reliable and sensitive tests for identifying the 'high-risk' foot.

AB - Aims To compare the Ipswich Touch Test and the VibraTip with the Neuropathy Disability Score and the vibration perception threshold for detecting the 'at-risk' foot. Methods We directly compared the Ipswich Touch Test and the VibraTip with both the Neuropathy Disability Score ≥6 and the vibration perception threshold ≥25V indicating 'at-risk' feet in 83 individuals. Results The vibration perception threshold and Neuropathy Disability Score tests exhibited almost perfect agreement with each other (P<0.001). The VibraTip and Ipswich Touch Test results were identical (P<0.001). The VibraTip and Ipswich Touch Test results also exhibited almost perfect agreement with the vibration perception threshold (P<0.001) and the Neuropathy Disability Score (P<0.001). Conclusions These two simple and efficient tests are easy to teach, reliable and can be used in any setting, and neither requires an external power source. We conclude that both the VibraTip and the Ipswich Touch Test are reliable and sensitive tests for identifying the 'high-risk' foot.

UR - http://www.scopus.com/inward/record.url?scp=84869144842&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84869144842&partnerID=8YFLogxK

U2 - 10.1111/j.1464-5491.2012.03730.x

DO - 10.1111/j.1464-5491.2012.03730.x

M3 - Article

VL - 29

SP - 1550

EP - 1552

JO - Diabetic Medicine

JF - Diabetic Medicine

SN - 0742-3071

IS - 12

ER -